Calibration, h, S sensor – RKI Instruments 35-3001-09 User Manual

Page 24

Advertising
background image

20 • 35-3001-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector

Returning to Normal Operation

1.

Wait 1 to 2 minutes to allow the oxygen reading at the controller to return to normal, then
return the controller to normal operation.

NOTE: If you do not allow the oxygen reading to return to normal, then unwanted alarms may

occur.

2.

Verify that the controller display reading stabilizes at 20.9%.

3.

Store the components of the calibration kit in a safe and convenient place.

Calibration, H

2

S Sensor

This section describes how to calibrate the hydrogen sulfide sensor of the sample-draw detector. It
includes procedures to prepare for calibration, set the zero reading, set the response reading, and
return to normal operation.

NOTE: This procedure describes calibration using a demand flow regulator.

Preparing for Calibration

1.

Follow the instructions in the controller’s operator’s manual for entering calibration mode.

NOTE: If you can verify a fresh air environment, it is not necessary to use the zero air calibration

cylinder to set the zero reading.

2.

Screw the regulator into a zero air calibration cylinder.

3.

Set a voltmeter to measure in the millivolt (mV) range and plug the leads into the H

2

S

amplifier.

Plug the positive lead into the red (+) amplifier test point; plug the negative lead into the black
(-) amplifier test point.

4.

Use the following formula to determine the correct test points output for the test
sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a test sample of 25 ppm and a fullscale setting of 100 ppm, the
correct output is 200 mV.

200 (mV) = (25/100) X 400 +100

Setting the Zero Reading

1.

Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet
line. This step is not necessary if you verified a fresh air environment earlier in this
procedure.

2.

Allow the sample-draw detector to draw sample for one minute.

3.

Verify a voltmeter reading of 100 mV (± 2 mV).

4.

If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the
voltmeter reading is 100 mV (± 2 mV). If you used a zero air calibration cylinder to set the
zero reading, proceed to step 5. If you verified a fresh air environment, proceed to the next

Advertising