RKI Instruments 35-3001-09 User Manual

Page 25

Advertising
background image

35-3001-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector • 21

section, Setting the Response Reading.

5.

Disconnect the sample tubing from the inlet line.

6.

Unscrew the regulator from the zero air calibration cylinder.

Setting the Response Reading

1.

Screw the regulator into the hydrogen sulfide calibration cylinder.

2.

Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet
line.

3.

Allow the sample-draw detector to draw sample for one minute.

4.

When the reading on the voltmeter stabilizes, verify that the reading matches the response
reading (±2 mV) you determined earlier.

5.

If necessary, use the span potentiometer on the amplifier to adjust the reading to match the
correct response reading.

6.

Disconnect the sample tubing from the inlet line.

7.

Unscrew the regulator from the hydrogen sulfide calibration cylinder.

Returning to Normal Operation

1.

Wait approximately one minute to allow the hydrogen sulfide reading to stabilize.

2.

Remove the voltmeter leads from the amplifier test points.

3.

Close the housing door.

4.

Follow the instructions in the controller’s operator’s manual to exit the calibration mode.

5.

Store the components of the calibration kit in a safe and convenient place.

Advertising