RKI Instruments 35-3001A-01H User Manual

Page 19

Advertising
background image

35-3001A-01H Hydrogen Sample-Draw Detector • 15

4.

Calibrate the sample-draw detector as described in “Calibration” on page 18.

5.

If the fail condition continues, replace the sensor as described in “Replacing the Hydrogen
Sensor” on page 15.

6.

If the fail condition continues, contact RKI Instruments, Inc. for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

The sensor responds slowly or does not respond during the monthly response test.

Unable to accurately set the zero or response reading during the calibration procedure.

The sensor requires frequent calibration.

Probable causes

The calibration cylinder is low, out-dated, or defective.

If a demand flow regulator calibration kit is used, the demand flow regulator is not functioning
properly.

The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed
pump, etc.

The sample-draw detector is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If a demand flow regulator calibration kit is used, use a different demand flow regulator to
determine if the original one is functioning properly.

3.

If necessary, set the correct flow rate with the sensor flow control valve.

4.

If you cannot set the correct flow rate, check the sample line for obstructions or kinks.

5.

If the calibration/response difficulties continue, replace the sensor as described later in this
section.

6.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further
instruction.

Replacing Components of the Hydrogen
Sample-Draw Detector

This section includes procedures to replace the sensor, hydrophobic filter, and particle filter.

Replacing the Hydrogen Sensor

1.

Turn off the controller.

2.

Turn off power to the controller.

3.

Open the housing door of the sample-draw detector.

4.

Unscrew and remove the two screws that secure the sensor retaining plate, then lift the plate,
connector, and sensor out of the housing.

5.

Unplug the connector from the sensor.

6.

Verify that you are using the correct replacement sensor (NC-6244A is printed on the sensor),
then plug the sensor into the connector.

7.

Place the sensor in the hydrogen sensor cavity, then position the retaining plate on the two
standoffs.

8.

Secure the retaining plate to the standoffs with the two screws you removed in step 4.

Advertising