16 • 35-3001A-01H Hydrogen Sample-Draw Detector

9.

Turn on power to the controller and place it into normal operation.

10. Turn on the controller.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue.

11. Calibrate the replacement sensor as described in “Calibration” on page 18.

Replacing the Amplifier

1.

Turn off the controller.

2.

Turn off power to the controller.

3.

Open the housing door of the sample-draw detector.

4.

Remove the detector and controller plug-in terminal strips by grasping the sides of the strips
with your fingers. Let the terminal strips hang by their connected wires while you replace the
amplifier.

5.

Unscrew the screw in the lower left corner of the amplifier and remove the screw, lock washer,
and flat washer. Be careful not to lose any of these parts.

6.

Remove the old amplifier from the main circuit board.

7.

Install the new amplifier in the same orientation as the old amplifier. See Figure 1.

8.

Reinstall the screw, lock washer, and flat washer you removed in step 5.

9.

Install the detector and controller plug-in terminals strips into their sockets on the new
amplifier. If controller leads or detector cable leads were removed from the plug-in terminal
strips during this procedure, refer to Table 2 and Table 3 below.

Table 2:Reconnecting the Amplifier
to the Detector/Amp Terminal Strip

Amplifier Controller

Terminal Strip

Detector/Amp Terminal

Strip on Main PCB

PWR/SIG “-”

-

PWR/SIG “S”

S

PWR/SIG “+”

+

Table 3:Reconnecting the LEL

Detector to the Amplifier

Amplifier Detector

Terminal Strip

Detector Lead

DETECTOR “R”

RED

DETECTOR “W”

WHT

DETECTOR “G”

GREEN

DETECTOR “B”

BLK