Calibration – RKI Instruments 35-3001A-05-02 User Manual

Page 23

Advertising
background image

35-3001A-05-02 Carbon Dioxide Sample-Draw Detector • 19

Calibration

This section describes how to calibrate the carbon dioxide sample-draw detector. It includes
procedures to prepare for calibration, set the zero reading, set the response reading, and return to
normal operation.

NOTE: This procedure describes calibration using a demand flow regulator.

Preparing for Calibration

1.

Follow the instructions in the controller’s operator’s manual for entering calibration mode.

2.

Open the housing door.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Plug the positive lead into the red (+) amplifier test point; plug the negative lead into the black
(-) amplifier test point labeled.

5.

Use the following formula to determine the correct test points output for the calibrating
sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 2,000 ppm and a fullscale setting of 5,000 ppm, the
correct output is 260 mV.

260 (mV) = (2000/5000) X 400 +100

Setting the Zero Reading

Since there is a background of CO

2

in air of typically 300 - 600 ppm, it is necessary to use a

calibration kit with a 100% nitrogen cylinder to set the zero signal of the carbon dioxide sample-
draw detector.

1.

Screw the regulator into a 100% nitrogen calibration cylinder.

2.

Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet
line.

3.

Allow the sample-draw detector to draw sample for one minute.

4.

Verify a voltmeter reading of 100 mV (± 2 mV).

5.

If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the
voltmeter reading is 100 mV (± 2 mV).

6.

Disconnect the sample tubing from the inlet line.

7.

Unscrew the regulator from the 100% nitrogen calibration cylinder.

Setting the Response Reading

1.

Screw the regulator into the carbon dioxide calibration cylinder.

2.

Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet
line.

3.

Allow the sample-draw detector to draw sample for one minute.

4.

When the reading on the voltmeter stabilizes, verify that the reading matches the response
reading (±2 mV) you determined earlier.

5.

If necessary, use the span potentiometer on the amplifier to adjust the reading to match the
correct response reading.

Advertising