20 • 35-3001A-05-02 Carbon Dioxide Sample-Draw Detector

6.

Disconnect the sample tubing from the inlet line.

7.

Unscrew the regulator from the carbon dioxide calibration cylinder.

Returning to Normal Operation

1.

Wait approximately one minute to allow the carbon dioxide reading to stabilize.

2.

Remove the voltmeter leads from the amplifier test points.

3.

Close the housing door.

4.

Follow the instructions in the controller’s operator’s manual to exit the calibration mode.

5.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 2 lists replacement parts and accessories for the carbon dioxide sample-draw detector.

Table 4: Parts List

Part Number

Description

06-1248RK

Sample tubing, 3/16 in. x 5/16 in., specify length

06-1248RK-03

Sample tubing, 3/16 in. x 5/16 in., 3 feet (for calibration kit)

07-0110RK

Gasket for IR flow block

30-1016RK

Pump

33-0165RK

Hydrophobic filter

33-0167RK

Particle filter

57-1053RK-10

Amplifier

61-5040RK-02

IR CO

2

sensor, 0 - 5,000 ppm

71-0340

Operator’s Manual, 35-3001A-05-02 Sample-Draw Detector

81-0070RK-01

Steel calibration cylinder, 2000 ppm CO

2

, 34-liter

81-0070RK-03

Steel calibration cylinder, 2000 ppm CO

2

, 103-liter

81-0078RK-01

Steel calibration cylinder, 100% nitrogen, 34-liter

81-0078RK-03

Steel calibration cylinder, 100% nitrogen, 103-liter

81-1054RK

Regulator, demand flow, for 58- and 103-liter cylinders

81-1055RK

Regulator, demand flow, for 17- and 34-liter steel cylinders