Calibration, oxygen detector – RKI Instruments 35-3010RKA-01 User Manual

Page 26

Advertising
background image

22 • 35-3010RKA-01 Sample-Draw Detector

5.

Store the components of the calibration kit in a safe and convenient place.

Calibration, Oxygen Detector

This section describes how to calibrate the oxygen detector in the sample-draw detector. It
includes procedures to prepare for calibration, set the zero reading, set the fresh air
reading, and return to normal operation.

NOTE:

This procedure describes calibration using a demand flow regulator, a zero air
calibration cylinder, and a 4-gas calibration cylinder to set the zero reading. A 4-
gas cylinder is used to set the zero since it is alreayd usd to calibrate the LEL, CO,
and H

2

S channels. A 100% nitrogen calibration cylinder could also be used to set

the zero reading.

Preparing for Calibration

CAUTION:

This procedure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.

1.

Open the housing door.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

For the oxygen channel, plug the voltmeter into the test points on the oxygen
transmitter. Plug the positive lead into the test point labeled TP+; plug the negative
lead into the test point labeled TP-.

4.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 12% and a fullscale setting of 25%, the
correct output for the oxygen test points is 292 mV.

292 (mV) = (12/25) X 400 +100

Setting the Zero Reading

NOTE:

The 4-gas calibration cylinder included in the calibration kit may be used to zero
the oxygen channel. Keep in mind that the oxygen concentration in that cylinder
is 12%, not 0%. Be sure to set the controller appropriately.

1.

Screw the regulator into the 4-gas calibration cylinder.

2.

Connect the calibration kit sample tubing to the regulator.

3.

Allow the sample-draw detector to respond to the calibrating sample for
approximately 1 minute.

4.

After one minute, check the mV output on the oxygen transmitter test points and
verify that the reading matches the response reading (±2 mV) you determined earlier
(292 mV for 12% oxygen).

5.

If necessary, use the span pot on the transmitter to adjust the reading to match the
correct response reading.

6.

Disconnect the sample tubing from the inlet line.