35-3010RKA-01 Sample-Draw Detector • 25

response reading.

8.

Disconnect the sample tubing from the sample draw detector’s inlet line.

9.

Unscrew the regulator from the calibration cylinder.

NOTE:

For convenience, leave the regulator connected to the sample tubing.

Returning to Normal Operation

1.

Remove the voltmeter leads from the test points.

2.

Reconnect the incoming sample line.

3.

Wait 1 to 2 minutes to allow the calibration gas to be drawn out and the readings to
stabilize.

4.

Close the housing door.

5.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 4 lists replacement parts and accessories for the sample-draw gas detector.

Table 2: Parts List

Part Number

Description

06-1248RK

Sample tubing, 3/16 x 5/16, specify length, (for calibration kit)

07-0034RK

Sealing gasket, for CO and H

2

S flow block cavities

30-0610RK

Pump

33-0171RK

Hydrophobic filter (AcroPak)

33-6095RK

Charcoal filter, CF-188

65-0601RK

Oxygen sensor

81-0076RK-01

Zero air calibration cylinder (34 liter)

81-0154RK-04

4-gas calibration gas cylinder, 50% LEL methane/12% oxygen/25
ppm H

2

S/50 ppm CO, 34 liter

81-1055RK

Regulator, demand flow, for 17 and 34 liter steel cylinders

ES-1531-CO

CO sensor

ES-1537-H2S

H

2

S sensor

NC-6245

Combustible sensor