Troubleshooting – RKI Instruments 61-0191RK User Manual

Page 13

Advertising
background image

61-1004RK/61-0191RK Carbon Dioxide Detector • 9

NOTE:

If the reading is not within ± 20% of the gas concentration, calibrate the detector
as described in “Calibration” on page 11.

9.

Turn the on/off knob clockwise to close the regulator.

10. Unscrew the regulator from the calibration cylinder and remove the sample tubing

from the regulator.

11. Unscrew the calibration cup from the detector. You may leave the sample tubing

connected to the calibration cup for convenience.

12. When the display reading falls below the alarm setpoints, return the controller to

normal operation or enable external alarms.

13. Store the components of the calibration kit in a safe place.

Biannually

Calibrate the CO

2

detector every six months as described in “Calibration” on page 11.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO

2

detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Table 3:Troubleshooting the CO

2

Detector

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The detector wiring is

disconnected or
misconnected.

• The detector zero

signal is low enough to
cause a fail condition.

• The detector is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Calibrate the detector.
3. If the fail condition continues, replace

the detector.

4. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Detector requires

frequent calibration.

Note: Under “normal”
circumstances, the
detector requires
calibration once every
six months.
Some applications may
require a more frequent
calibration schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The regulator flow rate

is not 0.5 LPM.

• The detector is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Confirm that the regulator being used

is a 0.5 LPM regulator.

3. If the calibration/response difficulties

continue, replace the detector.

4. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising
This manual is related to the following products: