Overview, Specifications – RKI Instruments 61-0191RK User Manual
Page 5

61-1004RK/61-0191RK Carbon Dioxide Detector • 1
Overview
This manual describes the 61-1004RK CO
2
(carbon dioxide) detector. This manual also
describes how to install, start up, maintain, and calibrate the detector when used with a
gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the CO
2
detector.
The 61-1004RK CO
2
detector includes the 61-0191RK detector and a junction box. This
manual may also be used for the 61-0191RK CO
2
detector which does not include a
junction box and is normally mounted in one of a controller’s conduit hubs and factory
wired to the controller. If you are using a 61-0191RK CO
2
detector, disregard all references
to the junction box and junction box terminal block.
Specifications
Table 1 lists specifications for the CO
2
detector.
WARNING: When using the 61-1004RK, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 61-1004RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 61-1004RK as described in this manual.
Table 1: 61-1004RK Specifications
Detection Range
With Junction Box
61-1004RK-02: 0 - 5,000 ppm CO
2
61-1004RK-03: 0 - 5% Volume CO
2
61-1004RK-05: 0 - 50% Volume CO
2
61-1004RK-10: 0 - 100% Volume CO
2
Without Junction Box
61-0191RK-02: 0 - 5,000 ppm CO
2
61-0191RK-03: 0 - 5% Volume CO
2
61-0191RK-05: 0 - 50% Volume CO
2
61-0191RK-10: 0 - 100% Volume CO
2
Area Classification
Explosionproof for Class I, Groups B, C, and D
Sampling Method
Diffusion
Response Time
90% in 45 seconds
Accuracy
± 5% of reading or ± 2% of full scale (whichever is greater)