Troubleshooting – RKI Instruments 61-0198RK User Manual

Page 13

Advertising
background image

61-1007RK/61-0198RK Carbon Dioxide Detector • 9

NOTE:

If the reading is not within ± 20% of the gas concentration, calibrate the detector
as described in “Calibration” on page 12.

9.

Turn the on/off knob clockwise to close the regulator.

10. Unscrew the regulator from the calibration cylinder and remove the sample tubing

from the regulator.

11. Unscrew the calibration cup from the detector. You may leave the sample tubing

connected to the calibration cup for convenience.

12. When the display reading falls below the alarm setpoints, return the controller to

normal operation or enable external alarms.

13. Store the components of the calibration kit in a safe place.

Biannually

Calibrate the CO

2

detector every six months as described in “Calibration” on page 12.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO

2

detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Table 2:Troubleshooting the CO

2

Detector

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The detector wiring is

disconnected or
misconnected.

• The plug-in sensor is

not properly plugged
into the three-socket
pattern in the detector
housing body.

• The detector zero

signal is low enough to
cause a fail condition.

• The detector is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Confirm that the plug-in sensor is

installed properly.

3. Calibrate the detector.
4. If the fail condition continues, replace

the detector.

5. If the fail condition continues, contact

RKI for further instruction.

Advertising
This manual is related to the following products: