Replacing the combustible detector – RKI Instruments 61-1000RKSS-05 User Manual

Page 13

Advertising
background image

61-1000RKSS-05 Combustible Gas Detector• 9

Replacing the Combustible Detector

1.

Turn off the controller.

2.

Turn off power to the controller.

3.

Remove the junction box cover.

4.

Disconnect the detector leads from the terminal block in the junction box. Note the
position of the color-coded leads as you remove them.

5.

Unscrew the detector from the junction box conduit hub.

6.

Guide the detector leads of the replacement detector through the junction box conduit
hub, then screw the mounting threads of the detector into the hub. If necessary for
environmental conditions, apply thread sealant or teflon tape to the hub and/or
detector threads to seal them.

7.

Connect the detector leads to the terminal block the same way the old detector was
wired (see Figure 3). See the controller operator’s manual and the controller’s detector
head specification sheet for the 61-1000RKSS-05 detector to verify the connections to
the controller are correct.

8.

Reinstall the junction box cover.

9.

Turn on or plug in power to the controller.

10. Turn on the controller and place into normal operation.

CAUTION:

Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.

11. Calibrate the replacement detector as described in the Calibration section of this

manual.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Detector requires

frequent calibration.

Note: Under “normal”
circumstances, the
detector requires
calibration once every
three months.
Some applications may
require a more frequent
calibration schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas is

not an appropriate
concentration.

• The detector is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the calibration gas

concentration is appropriate for the
detector.

3. If the calibration/response difficulties

continue, replace the detector.

4. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Table 2:Troubleshooting the Combustible Gas Detector (Continued)

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising