Calibration frequency, Calibration – RKI Instruments 61-1007RKSS User Manual

Page 16

Advertising
background image

12 • 61-1007RKSS Carbon Dioxide Gas Detector

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most infrared CO

2

detector

applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 6 months for the infrared
CO

2

detector.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment where CO

2

is not normally present, and calibration adjustments

are minimal at calibration, then a calibration frequency of every 9 to 12 months is
adequate.

If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 6 months may be necessary.

Calibration

This section describes how to calibrate the CO

2

detector. It includes procedures to prepare

for calibration, set the zero reading, set the response reading, and return to normal
operation.

WARNING: The controller is not an active gas monitoring device during the calibration

procedure.

The following procedure assumes the use of a calibration kit which includes a 100%
nitrogen calibration gas cylinder for setting the zero reading, a CO

2

calibration gas

cylinder for setting the response reading, a 0.5 LPM fixed flow regulator with an on/off
knob, a calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.

Preparing For Calibration

1.

Screw the calibration cup onto the detector.

2.

Use the sample tubing to connect the regulator to the calibration cup.

3.

Place the controller into its calibration program or disable external alarms.

NOTE:

Calibrating the CO

2

detector may cause alarms. Be sure to put the controller into

its calibration program or disable external alarms before continuing.

Setting the Zero Reading

Since there is a background of CO

2

in air of typically 300 - 600 ppm (0.03 - 0.06 %volume),

it is necessary to use a calibration kit with a CO

2

free gas to set the zero signal of a low

range CO

2

detector. Fresh air can be used to zero the detector if a 0-50 %volume or 0-100

%volume detector is being used.

1.

Follow the directions in the controller’s operator’s manual for setting the zero reading
(sometimes called the fresh air reading).

2.

When the directions call for exposing the detector to zeroing gas, screw the regulator
into the cylinder, turn the on/off knob counterclockwise to open the regulator, and

Advertising