Description, Oxygen detector, Amplifier detector terminal strip – RKI Instruments 65-2320RK-100 User Manual

Page 6: Figure 1: oxygen transmitter component location

Advertising
background image

2 • 65-2320RK-100 Oxygen Transmitter

Description

This section describes the components of the oxygen transmitter. The oxygen transmitter
is a 4 - 20 mA type detector head. It consists of the oxygen detector, amplifier, and junction
box.

Figure 1: Oxygen Transmitter Component Location

Oxygen Detector

The sensing components of the detector are encapsulated within a conduit mounting
black anodized aluminum housing. Through a series of chemical and electronic reactions,
the detector produces a millivolt output that is proportional to the detection range. 3/4’’
NPT mounting threads at the top of the detector allow you to mount the detector to the
junction box or a 3/4’’ NPT conduit fitting. Two color-coded leads extend from the top of
the detector. The leads allow you to connect the detector to a controller.

Controller Terminal Strip

- Test Point (Black)

100mV - 500mV

Range

T

O

X

IC

OX

Y

SI

G

/PW

R

S

Amplifier

Detector Terminal Strip

ZE

R

O

SP

A

N

Zero Pot

Factory Adjust Pot

+ Test Point (Red)

Span Pot

J-Box

Oxygen Detector

Advertising