Installation, Mounting the oxygen transmitter – RKI Instruments 65-2320RK-100 User Manual

Page 8

Advertising
background image

4 • 65-2320RK-100 Oxygen Transmitter

from a mounting surface and ensure that there is enough room to install a calibration cup
on the detector during calibration.

Installation

This section describes procedures to mount the oxygen transmitter in the monitoring
environment and wire the transmitter to a controller.

Mounting the Oxygen Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site that is at normal breathing level.

Figure 2: Mounting the Oxygen Transmitter

2.

At the monitoring site you select, hang or mount the junction box with the detector
facing down (see Figure 2).

.38

.75

3.65

3/4 NPT
Female

1 1/2-20 Thread
for Calibration Cup

Oxygen

Detector

J-Box

2.74

.10

Rubber
Spacer, 3X

7.9 MAX