65-2321RK Oxygen Transmitter • 5

from a mounting surface and insure that there is enough room to install a calibration cup
on the detector during calibration.

Installation

This section describes procedures to mount the oxygen transmitter in the monitoring
environment and wire the transmitter to a controller.

Mounting the Oxygen Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site that is at normal breathing level.

Figure 3: Mounting the Oxygen Transmitter

2.

At the monitoring site you select, hang or mount the junction box with the detector
facing down (see Figure 3).

J-Box

3.10

.38

.75

3/4 NPT
Female

Rubber Spacers 3X

Oxygen
Detector

3.65

6.8 Max

1-1/2 - 20 Thread for Calibration Cup