Replacing components of the oxygen transmitter – RKI Instruments 65-2322RK User Manual

Page 15

Advertising
background image

65-2322RK Oxygen Transmitter • 11

Replacing Components of the Oxygen Transmitter

This section includes a procedure to replace the plug-in oxygen sensor, a procedure to
replace the entire oxygen detector assembly, and one to replace the amplifier. In most
cases, it is not necessary to replace the entire detector assembly.

Replacing The Plug-in Oxygen Sensor

CAUTION: The plug-in sensor contains electrolyte that is an irritant. Do not disassemble the

sensor when replacing it with a new one. If electrolyte comes in contact with your
skin, wash affected area thoroughly with soap and water.

1.

Turn off the controller.

2.

Turn off or unplug power to the controller.

3.

Unscrew the detector housing cap from the detector housing body. Make sure not to
lose the cap gasket.

4.

Unplug and remove the oxygen sensor.

5.

Carefully plug the replacement sensor into the socket pattern that is located in the
detector housing.

6.

Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.

7.

Turn on power to the controller.

8.

Turn on the controller and place into normal operation.

CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the

next step.

9.

Calibrate the detector as described in “Calibration” on page 13.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the fresh air or
zero reading during
calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The flame arrestor in

the detector housing
cap is wet or clogged
with dirt or other
particulates.

• The calibration gas is

not an appropriate
concentration

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Check the detector housing cap to

determine if the flame arrestor is wet
or dirty. Clean or replace if necessary.

3. Verify that the calibration gas

concentration is appropriate for the
transmitter. Zero emission air (20.9%
oxygen) is normally used for a fresh
air adjustment if the environment is
suspect and 100% nitrogen (0%
oxygen) is normally used for a zero
adjustment.

4. If the calibration/response difficulties

continue, replace the plug-in sensor
as described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Table 2:Troubleshooting the Oxygen Transmitter (cont.)

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising