Calibration frequency, Calibration – RKI Instruments 65-2336RKSS User Manual

Page 18

Advertising
background image

14 • 65-2336RKSS CO Transmitter

amplifier as shown in Figure 4. If controller leads or detector leads were removed
during this procedure, refer to Figure 4 for the detector and amplifier connections.

NOTE:

When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be
sure to make arrangements so that this does not cause unwanted alarms.

10. Turn on or plug in power to the controller.

11. Turn on the controller and place it into normal operation.

12. Allow the transmitter to warm up for 5 minutes.

13. Calibrate the CO transmitter as described in “Calibration” on page 14.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most CO transmitter applications.
Unless experience in a particular application dictates otherwise, RKI Instruments, Inc.
recommends a calibration frequency of every 3 months for the CO transmitter.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.

If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 3 months may be necessary.

Calibration

This section describes how to calibrate the CO transmitter. It includes procedures to
prepare for calibration, set the zero reading, set the response reading, and return to
normal operation. It describes the test using a calibration kit that includes a calibration
cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI
Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.

Preparing for Calibration

NOTE:

Calibrating the CO transmitter may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before calibrating.

WARNING: Do not remove the detector housing cap or junction box cover while the

circuits are energized unless the area is determined to be non-hazardous.
Keep the detector housing cap and junction box cover tightly closed during
operation.

1.

Unscrew and remove the junction box cover.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.

4.

Use the following formula to determine the correct test points output for the

Advertising