Calibration frequency, Calibration – RKI Instruments 65-2391RK User Manual

Page 17

Advertising
background image

65-2391RK CO

2

Transmitter • 13

NOTE:

When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stabilizing and may be above the controller alarm
points or well below zero momentarily. RKI controllers have a one minute
warmup period when the controller does not display any gas reading or give any
alarm indication. The CO

2

transmitter’s 4-20 mA signal should be stable by the

time the controller’s warmup period is over.

CAUTION:

Allow the detector to warm up for 5 minutes before you continue with the next step.

14. Calibrate the CO

2

transmitter as described in “Calibration” on page 13 of this manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most infrared CO

2

transmitter

applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 6 months for the infrared
CO

2

transmitter.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment where calibration adjustments are minimal at calibration, then a
calibration frequency of every 9 to 12 months is adequate.

If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 6 months may be necessary.

Calibration

This section describes how to calibrate the CO

2

transmitter. It includes procedures to

prepare for calibration, set the zero reading, set the response reading, and return to
normal operation.

WARNING: The controller is not an active gas monitoring device during the calibration

procedure.

NOTE:

The following procedure assumes the use of a calibration kit which includes a
CO

2

calibration gas cylinder, a 100% nitrogen cylinder, a 0.5 LPM fixed flow

regulator with an on/off knob, a calibration cup for the detector, and a short
piece of sample tubing to connect the regulator to the calibration cup.

Preparing for Calibration

1.

Screw the calibration cup onto the bottom of the CO

2

detector.

2.

Use the sample tubing that comes with the calibration kit to connect the regulator to
the calibration cup.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Remove the junction box cover, then plug the voltmeter leads into the test points on

Advertising