Troubleshooting, Replacing components of the co, Transmitter – RKI Instruments 65-2396RK User Manual

Page 15

Advertising
background image

65-2396RK CO

2

Transmitter • 11

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO

2

transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Replacing Components of the CO

2

Transmitter

This section includes procedures to replace the CO

2

detector and amplifier.

Replacing the IR CO

2

Detector

1.

Turn off the controller.

2.

Turn off or unplug power to the controller.

3.

Remove the junction box cover.

4.

Remove the detector terminal strip from its socket.

5.

Disconnect the detector leads from the detector terminal strip. Note the position of the
color-coded leads as you remove them.

6.

Unscrew the detector from the junction box.

7.

Guide the detector leads of the replacement detector through the bottom conduit hub

Table 3:Troubleshooting the CO

2

Transmitter

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The transmitter wiring

is disconnected or
misconnected.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the transmitter wiring is

correct and secure.

2. Calibrate the transmitter.
3. If the fail condition continues, replace

the detector.

4. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 6
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas

flow rate is too low.

• The calibration gas is

not an appropriate
concentration.

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the regulator used for

calibration is a 0.5 LPM regulator.

3. Verify that the calibration gas

concentration is appropriate for the
transmitter. The concentration should
be in the detection range, preferably
about half of the detection range.

4. If the calibration/response difficulties

continue, replace the detector.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising