Calibration – RKI Instruments 65-2405RKSS User Manual

Page 17

Advertising
background image

65-2405RKSS Combustible Gas Transmitter • 13

Calibration

This section describes how to calibrate the combustible gas transmitter. It includes
procedures to prepare for calibration, set the zero reading, set the response reading, and
return to normal operation.

WARNING: The controller is not an active gas monitoring device during the calibration

procedure.

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.

Preparing for Calibration

1.

Screw the calibration cup onto the bottom of the detector.

2.

Use the sample tubing to connect the regulator to the calibration cup.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Remove the junction box cover, then plug the voltmeter leads into the test points on
the amplifier.

Plug the positive lead into the red + test point; plug the negative lead into the black -
test point.

5.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 50 %LEL and a fullscale setting of
100 %LEL, the correct output is 300 mV.

300(mV) = (50/100) X 400 +100

6.

Place the controller into its calibration mode or disable external alarms.

NOTE:

Calibrating the combustible gas transmitter may cause alarms. Be sure to put the
controller into its calibration program or disable external alarms before
continuing.

Setting the Zero Reading

NOTE:

If you can verify that the combustible gas transmitter is in a fresh air
environment, you do not need to apply zero air to the detector before adjusting
the zero reading.

1.

Screw the regulator into the zero air calibration cylinder.

2.

Turn the regulator knob counterclockwise to open the regulator.

3.

Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If
necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).

Advertising