RKI Instruments 65-2422RK-05 User Manual

Page 16

Advertising
background image

12 • 65-2422RK-05 H

2

S Transmitter

7.

Carefully plug the replacement sensor into the socket pattern that is located in the top
section of the detector housing.

NOTE:

Match the sensor’s male pins with the four female sockets as you plug the sensor
into the socket.

8.

Screw the bottom section of the detector housing onto the top section.

9.

Turn on or plug in incoming power at the power source end.

10. Turn on the controller.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.

11. Calibrate the replacement sensor as described in the Calibration section of this

manual.

Replacing the Amplifier

1.

Turn off the controller.

2.

Turn off or unplug incoming power at the power source end.

3.

Remove the junction box cover.

4.

Disconnect the detector leads from the detector terminal strip.

5.

Unscrew and remove the two screws that secure the amplifier to the junction box.
The screws are at the top left and bottom right of the amplifier.

6.

Remove the amplifier.

7.

Place the new amplifier in the same position as the old amplifier.

8.

Use the two screws you removed in step 5 to secure the new amplifier to the junction
box.

9.

Verify that the jumper block is installed over the TOXIC selector of the amplifier type
selector as shown in Figure 3.

10. Reconnect the wiring from the controller to the interconnect terminal strip as shown

in Table 2 and Figure 3.

Table 2: Reconnecting the H

2

S Amplifier to a Controller

Amplifier

Interconnect Terminal Strip

Controller Transmitter

Terminal Strip (typical)

4/20 FB

4 -20 (FB)

24V +

+ V (11 - 30 VDC)

Advertising