RKI Instruments 65-2423RK-05 User Manual

Page 12

Advertising
background image

8 • 65-2423RK-05 H

2

S Detector

Quarterly

Calibrate the H

2

S detector as described in “Calibration” on page 11.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the H

2

S detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Fail Condition

Symptoms

The controller indicates a fail condition.

Probable causes

The detector wiring is disconnected or misconnected.

The detector’s zero reading is low enough to cause a fail condition.

The detector is malfunctioning.

Recommended action

Verify that the detector wiring is correct and secure.

Calibrate the detector.

If the fail condition continues, replace the H

2

S sensor as described later in this section.

If the fail condition continues, contact RKI for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

The detector responds slowly or does not respond during the monthly response test.

Unable to accurately set the zero or response reading during the calibration
procedure.

The detector requires frequent calibration.

NOTE:

Under “normal” circumstances, the detector requires calibration once every three
months. Some applications may require a more frequent calibration schedule.

Probable causes

The calibration cylinder is low, out-dated, or defective.

The detector is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If the calibration/response difficulties continue, replace the H

2

S sensor as described

later in this section.

3.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for
further instruction.