RKI Instruments 65-2423RK-05 User Manual

Page 12

Advertising
background image

8 • 65-2423RK-05 H

2

S Detector

Quarterly

Calibrate the H

2

S detector as described in “Calibration” on page 11.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the H

2

S detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Fail Condition

Symptoms

•

The controller indicates a fail condition.

Probable causes

•

The detector wiring is disconnected or misconnected.

•

The detector’s zero reading is low enough to cause a fail condition.

•

The detector is malfunctioning.

Recommended action

•

Verify that the detector wiring is correct and secure.

•

Calibrate the detector.

•

If the fail condition continues, replace the H

2

S sensor as described later in this section.

•

If the fail condition continues, contact RKI for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

•

The detector responds slowly or does not respond during the monthly response test.

•

Unable to accurately set the zero or response reading during the calibration
procedure.

•

The detector requires frequent calibration.

NOTE:

Under “normal” circumstances, the detector requires calibration once every three
months. Some applications may require a more frequent calibration schedule.

Probable causes

•

The calibration cylinder is low, out-dated, or defective.

•

The detector is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If the calibration/response difficulties continue, replace the H

2

S sensor as described

later in this section.

3.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for
further instruction.