Installation, Mounting the h, S detector – RKI Instruments 65-2423RK-05 User Manual

Page 7

Advertising
background image

65-2423RK-05 H

2

S Detector • 3

Installation

NOTE:

Normally, the H

2

S detector is shipped already installed in one of the bottom

conduit hubs of a controller and factory wired to the controller. If your detector is
already installed in a controller, use this section for reference only.

This section describes procedures to mount the H

2

S detector in the monitoring

environment and wire the detector to a controller.

Mounting the H

2

S Detector

1.

Whether you will mount the detector directly to a controller or to a junction box, select
a mounting site that is representative of the monitoring environment. Consider the
following when you select the mounting site.

Select a site where the detector is not likely to be bumped or disturbed. Make sure
there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site where the target gas is likely to be found first.

NOTE:

If your application does not require a specific mounting site, mount the detector
at approximately breathing level.

Figure 2: Outline & Mounting Dimensions

3/4 NPT

Shrink Tubing

1.62 Flats for
Wrench

2.25 Ref

1-1/2-20 For Calibration Cup

9.75 Ref

0.75

Ш 1.75

3.48