RKI Instruments 65-2480RK User Manual

Page 37

Advertising
background image

65-2480RK Multi Point Detector • 33

6.

Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.

7.

Turn on or plug in power to the controller.

8.

Turn on the controller and place into normal operation.

CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the

next step.

9.

Calibrate the detector as described in “Calibration, O

2

Detector” on page 39.

Replacing the Oxygen Detector

NOTE:

In most cases, it is only necessary to replace the oxygen sensor.

1.

Turn off the controller.

2.

Turn off or unplug incoming power to the controller.

3.

Remove the junction box cover.

4.

Disconnect the detector leads from the terminal block in the junction box. Note the
position of the color-coded leads as you remove them.

5.

Unscrew the detector from the junction box hub.

6.

Guide the detector leads of the replacement detector through the junction box hub
then screw the mounting threads of the detector into the hub. If necessary for
environmental conditions, apply thread sealant or teflon tape to the hub and/or
detector threads to seal them.

7.

Connect the detector leads to the terminal block the same way the old detector was
wired. See the controller operator’s manual and the controller’s detector head
specification sheet for the oxygen detector to verify the connections are correct. Also
see Figure 15-Figure 23 for wiring diagrams.

8.

Reinstall the junction box cover.

9.

Turn on or plug in power to the controller.

10. Turn on the controller.

CAUTION:

Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.

11. Calibrate the replacement detector as described in “Calibration, O

2

Detector” on

page 39.

Advertising