14 • 65-2484RK-06 Multi Point Detector

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the multi-point detector.

NOTE:

This troubleshooting guide describes multi-point detector problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Table 2: Troubleshooting the Multi-Point Detector

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

Controller indicates a
fail condition.

• The detector wiring is

disconnected or
misconnected.

• The O

2

, H

2

S, or SO

2

plug-in sensor is not
properly plugged into
the sockets in the
detector housing body.

• The detector’s zero

reading is low enough
to cause a fail
condition.

• The detector is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Confirm that the O

2

, H

2

S, or SO

2

plug-in sensor is installed properly.

3. Perform a zero adjustment. A full

calibration is recommended.

4. If the fail condition continues, replace

the O

2

, H

2

S, or SO

2

plug-in sensor as

described later in this section.

5. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Detector requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The flame arrestor on

the LEL detector or in
the O

2

, or H

2

S detector

housing cap is wet or
clogged with dirt or
other particulates.

• The calibration gas is

not an appropriate
concentration.

• The detector is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Check the LEL detector or detector

housing cap to determine if the flame
arrestor is wet or dirty. Check the
sensor face of the SO

2

plug-in sensor

to determine if it is wet or dirty. Clean
if necessary.

3. Verify that the calibration gas

concentration is appropriate for each
detector.

4. If the calibration/response difficulties

continue, replace the LEL detector or
the O

2

, H

2

S, or SO

2

plug-in sensor as

described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.