10 • 65-2495RK/65-2498RK H

2

S Detector

WARNING: You must replace the plug-in sensor with the same type of sensor that is

installed. A detector cannot be converted from one type of detector to
another by using a different plug-in sensor. For example, if you are
replacing an H

2

S sensor, you must replace it with an H

2

S sensor.

7.

Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.

8.

Turn on power to the controller.

9.

Turn on the controller.

CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the

next step.

10. Calibrate the detector as described in “Calibration” on page 12.

Replacing the Hydrophobic Membrane

1.

Turn off the controller.

2.

Turn off or unplug incoming power to the controller.

3.

Unscrew the detector housing cap from the detector housing body.

4.

Gently pry up the edge of the white hydrophobic membrane with a small flat blade
screwdriver or a similar tool.

5.

Peel off the hydrophobic membrane. It may be necessary to clean off the detector
housing cap face to remove any residue left from the adhesive backed membrane.

6.

Install the new membrane in the recess on the face of the detector housing cap.

7.

Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.

8.

Turn on power to the controller.

9.

Turn on the controller.

Replacing the H

2

S Detector

NOTE:

In most cases, it is only necessary to replace the plug-in sensor.

1.

Turn off the controller.

2.

Turn off or unplug incoming power to the controller.

3.

If the detector is installed directly on a controller, open the controller door.

If the detector is installed remotely from a controller in a junction box, remove the
junction box cover.

4.

If the detector is installed directly on a controller, disconnect the detector leads from
the detector terminal strip in the controller. Note the position of the color-coded leads
as you remove them.

If the detector is installed remotely from a controller in a junction box, disconnect the
detector leads from the terminal block in the junction box. Note the position of the
color-coded leads as you remove them.

5.

Unscrew the detector from the controller conduit hub or junction box conduit hub.