Internal description, Terminal pcb, View with cover removed – RKI Instruments 65-2610RK-04 User Manual

Page 10: View with cover and control pcb removed, Modbus terminal strip

Advertising
background image

5

65-2610RK-04 M2 Transmitter Operator’s Manual

Internal Description

This section describes the internal components of the M2. The internal components of the
M2 include the terminal PCB which provides for all the wiring connections to the M2 and
the control PCB which displays the gas reading and has the control buttons.

Figure 2: M2 Internal Components

Terminal PCB

The terminal PCB is encapsulated in epoxy for protection against moisture and physical
damage. It is mounted into the rear of the junction box with three standoffs and rests on a
thin layer of foam. A banana jack is screwed into each of the standoffs and used for
mounting the control PCB. The terminal PCB converts the electrical output from the
detector to a signal which can be displayed by the LCD display, a 4 - 20 mA signal (that is
proportional to the detection range), and an RS-485 Modbus output signal. The 4 - 20 mA
signal may be used by a recording device, gas monitor controller, or programmable
controller. The Modbus output may be used to connect the M2 to a Modbus network. The
terminal PCB also controls three relays, one fail and two gas alarm relays.

Two columns of plug-in style terminal strips are used to make all wiring connections to
the M2. The column on the left consists of the power/signal, detector, and Modbus
terminal strips. The column on the right consists of the relay terminal strips. A 20 position
connector at the bottom of the terminal PCB is used to connect the terminal PCB to the

Alarm 1 LED

Fail LED

TX LED

RX LED

Alarm 2 LED

UP/YES

ENTER

LCD Display

ENTER

DOWN/NO

Push Button

Control Switches

C

N

C

N

O

AL

AR

M

1

LE

L

R

W

G

B

R

W

G

B

FA

IL

C

N

C

N

O

PW

R

/SI

G

-

S

+

-

S

+

Detector Terminal Strip

Power/Signal

Terminal Strip

Programming

Connector

(Factory Use)

Detector

Current Pot

(Factory

Adjust)

UP/YES

DOWN/NO

Termination

Jumper

View With Cover Removed

Modbus

Terminal Strip

View With Cover and

Control PCB Removed

Display
Ribbon Cable
Connector

3/4" NPT Conduit Opening

for Wire Entry

LE

L

P

W

R

/S

IG

RKI INSTRUMENTS

M2 TRANSMITTER

A

B

C

R

S

485

A

B

C

RS

4

8

5

AL

A

R

M

2

C

N

C

N

O

Relay
Terminal
Strips

Junction Box

Magnetic
Switches

Terminal PCB

Control PCB

3/4" NPT Conduit
Opening for
Detector J-Box
Wiring