65-2615RK-05-04 M2 Transmitter Operator’s Manual

28

seconds before returning to Configuration Mode. Repeat steps 6 - 8 until the settings
are as desired.

9.

After the M2 completes its warm-up sequence, the display will indicate CAL
NEEDED

. Since the gas type has been changed, a successful calibration must be

performed before the M2 can enter normal operation.

10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must

hold the UP/YES button for 5 seconds to enter Calibration Mode, but since a
calibration is required, only a momentary push is needed to enter Calibration Mode
after changing the gas type.

11. See “Calibration” on page 33 for calibration instructions.

NOTE:

When calibrating an M2 after changing the gas type, the M2 will not ask if you
want to calibrate, or whether you want to perform a fresh air adjustment, span
adjustment or zero adjustment. Since a complete calibration is required, the
calibration sequence will proceed without asking if you want to perform
certain operations.