Specifications – RKI Instruments 65-2615RK-05-04 User Manual

Page 7

Advertising
background image

65-2615RK-05-04 M2 Transmitter Operator’s Manual

2

Specifications

Table 1 lists specifications for the M2.

WARNING:

When using the M2, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the M2 and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the M2 as described in this manual.

WARNING:

M2 detector heads with firmware version 5.0 or later have a different
Modbus register assignment than M2s with previous versions of
firmware. If you are adding M2s with firmware version 5.0 or later to an
existing Modbus network installation, confirm what firmware version is
installed in the M2s already in the network and if necessary, take all steps
required to address the differences in Modbus register assignments. See
“Supported Modbus Functions” on page 42 for information on the current
Modbus register assignments. The firmware version is shown on the
Information Screen described on page 18.

Table 1: M2 Specifications

Target Gas/Detection Range

Hydrogen Sulfide:
0 - 100 ppm, 1 ppm increments

Alarm Settings (Alarm 1/Alarm 2)

10 ppm/50 ppm*

Construction (housing)

Explosion-proof Junction Box, NEMA 4

Area Classification

Explosion-proof for Class I, Groups B, C, and D

Sampling Method

Diffusion

Input Power

19 - 30 VDC

Controls

• Three push button switches
• Three magnetic switches for non-intrusive calibration

Weight

5.5 lbs.

Signal Output

• 4 to 20 mA, 500 ohms impedance max
• RS-485 Modbus

Operating Temperature

-20°C to 50°C

Accuracy

± 5% of reading or ± 2 ppm H

2

S (whichever is greater)

*

These are the RKI factory settings. You can change the alarm settings in Configuration Mode. See

“Viewing & Changing M2 Parameters” on page 23.

Advertising