5 • Model GD-K71D4X Sample Draw Transmitter

CAUTION:

Leave the cable shield’s drain wire insulated and disconnected at the
GD-K71D4X. You will connect the opposite end of the cable shield’s drain wire at the
controller.

CAUTION:

At the controller, do not route power and GD-K71D4X wiring through the same
conduit hub. The power cable may disrupt the transmission of the GD-K71D’s
signal to the controller.

Start Up

See the GD-K71D Operator’s Manual for GD-K71D startup instructions.

Operation

See the GD-K71D Operator’s Manual for an operational description of the GD-K71D that
is mounted inside the enclosure.

Maintenance/Calibration

See the GD-K71D Operator’s Manual for maintenance and calibration instructions. Use
the inlet fitting on the NEMA 4X enclosure to apply gas instead of the fittings inside the
enclosure on the GD-K71D.

Parts List

Table 4 lists replacement parts and accessories for the GD-K71D.

Table 2: Parts List

Part Number

Description

06-1273RK

1/4” O.D. x 1/8” I.D. Teflon PTFE tubing

17-4820RK

Sample fitting, 1/4” tube bulkhead union

18-0107RK

3/4” conduit hub

33-0165RK-01

Hydrophobic filter w/ flexible tubing stubs