Refrigeration & lighting system, Alarm instructions, Resetting the unit – Summit ACR1515SS User Manual

Page 10

Advertising
background image

10

REFRIGERATION & LIGHTING SYSTEM

The compressor in your ACCUCOLD® refrigerator is designed to function silently. DO NOT plug in the
unit immediately after moving it or changing its position. Allow a minimum of 1 hour for the pressures
inside the refrigeration system to equalize and the oils in the compressor to settle. After plugging the unit
in, let it run for one more hour without load or product. The refrigerator may then be loaded.

The temperature is controlled by a thermostat which is located on the front ventilation louvers, just below
the door. The thermostat has been factory-set to keep the product between 35.6 and 46.4°F (2 and 8°C).
The thermostat normally displays the temperature inside the cabinet.

TEMPERATURE CONTROL

To change the set temperature, push the middle button on the thermostat until the previous set
temperature is displayed. Push the upper or lower button until the desired temperature value is shown.
Finally, push the middle button again to select the setting.

To find out about changing other parameters, call our Customer Service Line at 1-800-932-4267.

Cabinet lighting is designed to function continuously or may be controlled with a switch, located inside the
unit to the left of the internal fan in the ceiling. See Wiring Diagram, page 8. Fluorescent lighting surrounds
the glass on the interior side of the door. Additionally, the header panel above the door is backlit.

LIGHTING



ALARM INSTRUCTIONS


Use a pointed object to push the RESET button on the back of the alarm unit. It is
necessary to reset the unit whenever a change is made to °C/°F temperature or
FAST/NORMAL sampling rate, or if the probe sensor is plugged in or removed.

Resetting the Unit

Note: HI/LO temperature alarm settings and MIN/MAX memories will be cleared
after resetting the unit.

DO NOT TOUCH THE COMPRESSOR TOP COVER

while unit is running. It is designed for high
operational temperatures and its surface can
become very hot, which may cause burns.
(The compressor is located at the bottom of
the refrigerator, behind the front louvers.)

NOTES:
• If the unit is unplugged, or power is lost or turned off, you must wait 3 to 5 minutes before

restarting the unit. If you attempt to restart before this time delay, the unit will not start.

• When you use the unit for the first time or restart the unit after it has been shut off for a long

time, allow a few hours for the interior temperature to normalize.

Advertising