11. The light is now ready for use.


Only use the following bulbs:
Sunnex part# 18560
3x35W: Sylvania Tru-Aim Titan FRB 35MR16Q/10/NSP/T
3x35W: Osram IRC 35W 10°

CAUTION! A CHOICE OF A REPLACEMENT LAMP/REFLECTOR CAN
AFFECT THE PERFORMANCE OF YOUR TRISTAR LIGHT. WE
STRONGLY SUGGEST THAT ONLY OSRAM/SYLVANIA TRU-AIM TITAN
ESX LAMP/REFLECTORS BE USED. THESE

ARE DESIGNED TO FILTER BOTH INFRARED AND ULTRAVIOLET
RAYS, PROVIDING YOU A CONTINUED SAFE, COOL PERFORMANCE.
THEY PRODUCE A PATTERN WITH A TIGHT CENTER, HARMONIOUS
TO A TEN INCH PERIMETER. INCREASED FRONTAL HEAT AND A
DISTORTED PATTERN ARE BUT A COUPLE OF THE DISADVANTAGES
OF PURCHASING INCORRECT REPLACEMENT LAMP REFLECTORS.

8.2 FUSE REPLACEMENT

1. Remove the plug from wall receptacle.

2. Access the transformer housing and detach the front cover by

removing the four plastic screws using a flat screwdriver.

3. Locate fuse blocks (2) in the transformer.

4. Turn the fuse holder counter clockwise using a suitable tool and

remove the fuse holder from the fuse block by pulling it out.

5. Replace fuse with proper replacement fuse.
Fuse Replacement Specifications:

USA (110V): Buss Fuse – MDA 2 250 Volt OR Interpower 813MDA-2

(250V, 2 Amp Fuse, MUST USE A SLOW-BLOW FUSE)

Europe (230V): Littlefuse – T-1.25A-218 (250V, 1.25 Amp Fuse,

MUST USE A SLOW-BLOW FUSE)

6. Follow procedure in reverse to return lamp to operating condition.