background image

assembly with the new replacement assembly and insert and tighten the four

screws. T

o remove the tube, grasp the tube ends by twisting one quarter turn until

the tube is free.

Handheld Lamps ... T

o change the filter on six-watt models, the filter assembly

snaps out of the lamp housing. Remove the lower housing. T

o remove the tube,

carefully grasp the tube ends by twisting one quarter turn until the tube is free.

Insert the new tube, giving it a twist to lock in place. Reassemble filter assembly.

Note: when replacing a multi-band shortwave/longwave tube, be sure to place

the longwave tube side on the proper end.

REPLACEMENT PA

RTS

MODEL

TUBE/WAVELENGTH

FILTER/FRAME

FOR HANDHELD LAMP MODELS

UVG-54

34-0013-01 254nm shortwave

98-0004-03

UVGL-55

34-0015-01 254/365nm

98-0005-03

UVL-56

34-0034-01 365nm longwave

98-0006-03

UVGL-58

34-0013-01 254nm shortwave

98-0004-03

34-0034-01 365nm longwave

UVM-57

34-0044-01 302nm midrange

98-0004-03

UVL-13R

34-0082-01 365nm BL

98-0006-03

UVS-13

34-0081-01 254nm shortwave

98-0004-03

FOR COMPACT LAMP MODELS

UVG-11

34-0003-01 254nm shortwave

98-0016-02

UVGL-15

34-0004-01 254/365nm

98-0017-02 (SW)

UVL-21

34-0005-01 365nm longwave

UVGL-25

34-0003-01 254nm shortwave

98-0016-02

UVL-23

UVL-23R

UVL-23RW

UVL-9R

34-0005-01 365nm longwave

34-0010-01 365nm

longwave

34-00

10-01

365nm longwave

34-0005-01 365nm longwave

34-0080-02 365nm longwave

REPAIRS

Contact UVP’s offices with any questions. T

o return a unit to the factory for repair,

call UVP’s Customer Service Department for a Return Authorization Goods (RGA)

number prior to sending the unit. Contact:

Corporate Headquarters: UVP

, LLC, 2066 W.

11th Street, Upland, CA 91786

Tel: (800) 452-6788 · (909) 946-3197

European Operations: Ultra-Violet Products Ltd., Unit 1, Trinity Hall Farm Estate,

Nuffield Road, Cambridge CB4 1TG UK • Tel: +44(0) 1223 420022

USING THE ULTR

AVIOLET LAMP

Plug your lamp into a standard AC outlet. Single

ultraviolet wavelength lamps utilize an ON/OFF

rocker switch. Multiple ultraviolet wavelength

lamp models use an ON/OFF/ON rocker switch

where the ON corresponds to the UV wave-

lengths. These ultraviolet lamps are designed

for use in dark or semi-dark areas.

Allow enough time for your eyes to

adjust to the darkness prior to using

the lamp. Some materials will tend

to fluoresce brighter than others. This

reaction is due to the concentration

of fluorescence on the material and

varying degrees of brightness of dif-

ferent colors.

UVL-13R and UVL-9R operate with

a 12V connector which plugs into an

automobile cigarette lighter. These portable lamps are excellent for field use and

can be used for windshield curing. The lamp can be attached to the window with

suction cups (purchased separately). These lamps can also be operated from a

wall outlet or with the use of an optional battery pack and charger, or alligator clips

(see accessories for ordering information).

USING THE WA

VELENGTH SELECTOR (Models

UVGL-55, UVGL-15)

The wavelength selector is included on multi-band lamps — half longwave and

half shortwave. The wavelength selector is designed to block off either longwave

or shortwave on these lamps. The selector can be slid up and down to give you

either wavelength. When the selector is moved over the bottom half of the filter

area, it blocks the longwave UV

, allowing only shortwave UV to be emitted. By

sliding the selector over the top half of the filter, only the longwave UV is emitted.

The wavelength selector snaps on or off the lamp easily.

WARNING:

REMOVE POWER CORD FROM PO

WER SOURCE

PRIOR TO SERVICING

.

FILTER ASSEMB

LY AND TUBE REPLACEMENT

Filters on the longwave lamps last indefinitely and should never need replacing

unless broken. Lamps with a BLB tube do not use a filter as the BLB tube is self-

filtered. Shortwave and midrange filters have an average life of 1000 hours.

Compact Lamps ... T

o replace the filter or UV tube on the four-watt compact lamp

assembly, remove the product label on the face of the lamp as one (or two de-

pending on the model) of the screws is located under the label.

T o remove the filter and frame assembl

y, unscrew the four retaining screws.

Two

screws are located above the filter and one below

. Replace the filter and frame

Compact Lamp Style

Handheld Lamp Sty

le

Advertising
This manual is related to the following products: