BC-300 Benchtop UV Curing System

7

PROBLEM

CHECK

SOLUTION

House power on but no belt

F1 in control box

Replace fuse

movement or panel lights

5 Amp fuse

Control box and belt working but

F1 in rear of power supply

Replace fuse

no control power to power supply

5 Amp fuse

Lamp does not light

Lamp condition

Replace fuses or lamp

Main fuses

Primary voltage

Check all primary and secondary

Repair or replace wires or

wiring for burnt or loose connections

connectors

Check all capacitors

Replace if bad

Check high voltage cable in rear of

Replace if needed

power supply and in irradiator for

loose connections

Check for proper seating between

the male and female on top of

the irradiator

Conveyor motor will not start

Check DC output from SCR in

If AC input with no DC output,

control box terminal block A+, A-.

replace SCR

If no DC output, check all AC input

to SCR.

Check at motor for DC power. If

Replace motor

you have power remove motor from

chain and try with no load. If it does

not run, replace.

If it does move, check chain and

Repair as needed

drive for loose or broken parts

Vacuum blower will not run

Check fuses F2, F3 in control box

Replace if blown

With blower switch on, check for

Replace switch or receptacle

power at blower receptacle

as needed

If blower receptacle has power,

Replace motor if needed

check blower cord and connections.

If OK, replace motor.

Lamp comes on but secondary

Check lamp condition and hours

Replace lamp

ammeter does not drop to

proper reading after 5 minutes

Lamp will not come up to full

Check for bad capacitor. Check for

Replace if needed

power and start up

mercury relay stuck in open position

Lamp intensity will not change

With power off, check mercury relays. Replace relays if needed

using rotary switch

They are normally closed.