Weidmuller PMX400TMP - Display instrument for control panel User Manual

Page 8

Advertising
background image

1

Note: the analogue output calibration sequence is not shown if the
analogue output is disabled.
You can calibrate each stage in sequence or skip sections by pressing
the PGM button.

Whenthe

displayshows

Action/Description

Put the instrument in setup mode and scroll through the main menu

CALIB

Press

ENT to enter calibrate mode

ECALn

Press

ì or Å and then ENT to select millivolt input calibration (or PGM to skip)

E=0

Apply a 0.00 mV signal and press

ENT

E=10

Apply a 10.00 mV signal and press

ENT

E=20

Apply a 20.00 mV signal and press

ENT

E=40

Apply a 40.00 mV signal and press

ENT

E=80

Apply a 80.00 mV signal and press

ENT

E=200

Apply a 200.00 mV signal and press

ENT

jCALn

Press

ì or Å and then ENT to select CJC input calibration (or PGM to skip)

CJ=0

Apply a 0.00 mV signal to the CJC input and press

ENT

Cj=60

Apply a 600.0 mV signal and press

ENT

rCALn

Press

ì or Å and then ENT to select Rtd input calibration (or PGM to skip)

-200C

Set the RTD simulator to -200 ˚C and press

ENT

0C

Set the RTD simulator to 0 ˚C and press

ENT

400C

Set the RTD simulator to 400 ˚C and press

ENT

800C

Set the RTD simulator to 800 ˚C and press

ENT

OCALn

Press

ì or Å and then ENT to select analogue output calibration (or PGM to skip)

I=0

Measure the current (mA) output.
Adjust using the

ì and Å keys until the output is at 0.00 mA.

Press

ENT.

I=20

Adjust using the

ì and Å keys until the output is at 20.00 mA.

Press

ENT.

E=0

Measure the output voltage.
Adjust using the

ì and Å keys until the output is at 0.00 V.

Press

ENT.

E=10

Adjust using the

ì and Å keys until the output is at 10.00 V.

Press

ENT.

CALIB

You are now back in the main menu, press

PGM

SAvE

Instrument will save the new values and return to run mode