Inputcalibration, Instrumentsetup(continued) – Weidmuller PMX420SERIES - Display instrument for control panel User Manual

Page 24

Advertising
background image



Timer delay

dL1=

0

Timer delay (set to 0 s to disable)
e.g., no delay

-

Dec

-

Inc

Next

Accept

Calibrationoptions

Calibrate
Input?

CIPn

CIPy

Skip input calibration
Calibrate inputs

Toggle

Accept

Calibrate
Output?

COPn

COPy

Skip output calibration
Calibrate outputs

Toggle

Accept

Save values

SAVE

Instrument is saving the changes to the setup and returning to normal
operation

*Note: Alarms and Analogue outputs are fitted to the PMX420Plus only. You can still set values for the
alarms and analogue output on the PMX420 but they will only affect the display and status leds.

During setup the Å and ì keys change values and toggle options.

The ENT key accepts changes and moves to the next item in the setup
sequence (see table).
Note: if you remove the security link during the setup sequence, any
changes you have made will not be saved.

•InputCalibration

General
The PMX0 series instruments are factory calibrated. You can change
the input range without calibration, but the outputs are calibrated for
the specific range given during the setup procedure.

Equipmentrequirements
• A suitable current or voltage source

TerminalConnectionsforinputcalibration

CalibrationStage

Signaltype

Terminal

Current (mA) inputs

Milliamp source (+)

8

Milliamp source (−)

9

Voltage (V) inputs

Voltage source (+)

7

Voltage source (−)

9

InstrumentSetup(continued)