Before using this unit, Installation procedure – Yamato Scientific CLH600 Cool Line User Manual

Page 11

Advertising
background image

8

Before Using This Unit

Installation Procedure

6

Connecting the power.
Confirm that the leakage breaker and the power switch are turned "off," and then plug into an outlet.

7

Installing the external water tank (optional accessory)
Install the external water tank in a higher place than the unit. If it is installed in a lower place, the
flow rate may drop or air bleeding may not be carried out smoothly when pouring the circulating fluid
for the first time.

8

Pour the circulating fluid into the water tank.

• Remove the lid from the external water tank, and pour the circulating fluid.

(Select a circulating fluid in response to the set temperature condition.)

• Open the air release valve cock.
• Push the priming pump more than ten times.

(Feed the fluid from the external water tank to the circulating pump inside the unit to let out air
from the pump. The fluid circulates after air is let out.)

• Turn on the leakage breaker and the power switch. (The circulating fluid flows into the fluid tank

of the unit.)

• After the fluid tank is filled with the fluid, it is discharged into the external water tank. (If the

circulating fluid still does not circulate, immediately turn off the leakage breaker and the power
switch. Check the unit according to the procedure described on page 54.

• Close the air release valve cock.

The circulating pump may malfunction if the unit is operated with the circulating fluid
uncirculated.

• After the circulation of the circulating fluid is stabilized, resupply the circulating fluid to 80% level

of the external water tank.

• After the resupply of the circulating fluid is completed, turn "off" the leakage breaker.
• Replace the lid on the external water tank.

Exercise care not to allow the circulating fluid to get on the unit.If it gets on any electric
part, leakage or electric shock may result.If it splashes on the operation panel, wipe it out.