Operation method, Fixed temperature operation – Yamato Scientific CLH600 Cool Line User Manual

Page 20

Advertising
background image

17

Operation Method

Fixed Temperature Operation

In this mode, the unit starts to operate by pressing RUN/STOP key and continues operating at the
set temperature until RUN/STOP key is re-pressed, as shown in the figure below.

温度(℃)

時間(t)

SV:温度設定値

SV

スタート/ストップ キーON

スタート/ストップ キーOFF

Fixed temperature

operation procedure

1. Turn on the power (turn on the breaker in front)

™

Current version of the software is displayed for about four seconds

after turning on the power. The screen then displays the initial
setting. The current temperature in bath, operation mode
character and setting temperature of overheating prevention device
are displayed on respective screens.

2. Select the operation mode

• Press the FIXED TEMP key to display "FIX", which indicates

the fixed temperature operation, on the center display screen.

3. Set the temperature

Press the FIXED TEMP key again.

The setting temperature screen displays the character "Sv"
which indicates the temperature setting. Also it displays the
current setting temperature with blinking. The FIXED TEMP
lamp blinks, too.

Set the temperature by pressing the "▼▲".

The temperature can be set to the first decimal place.

Measurement temperature screen:
Displays the current temperature in bath.
Setting temperature screen:

Displays the operation mode character. (Refer to Page 13)

Overheating prevention screen:
Displays the setting temperature of overheating prevention device

TEMP

RUN/STOP key: ON

RUN/STOP key: OFF

TIME

SV: Set Temp.