Maintenance

Daily Maintenance and Inspection

„

Adjustment of Tension of V Belt

Various problems may result from improper tension of the V belt. To avoid these, check and adjust the
tension approximately once every six months.

Adjustment Procedure

(1) Make sure that you unplug the power cord before

adjustment.

(2) Detach the left and rear covers.

(3) Loosen the four bolts that fix the motor mount

plate. (Loosen them only--you do not have to

remove them.)

(4) Loosen the lock nut and turn the adjuster bolt to adjust the tension of the belt so that the belt deflects by 8

mm when it is pulled by a 2 kg(f) weight.

(5) After adjustment, tighten the lock nut and then tighten the four fixing bolts.

(6) The following conditions may be observed if the belt is improperly tightened.

If the belt is too loose:

1. Slip (speed does not increase) or heating of the pulley
2. Wearing or cracking of the V belt
3. Unusual noises

If the belt is too tight:

1. Abnormal heating (

110°C) of the motor (normally

80°C)

2. Damage to the motor shaft

(7) If the V belt wears severely, replace it with a new one. The model number of the V belt is M-40.

Shaker motor

V belt

Motor mount plate

Fixing bolts

V belt

Adjuster bolt

Lock nut

V belt

Large pulley

Small pulley