Operation Method
Calibration Offset Function
3
Set calibration offset
TROUBLE
ENTER
CANCEL
Calibrate
+2.5℃
(Ex.)
Indicated temperature on controller: 37℃
Measured temperature: 36℃
↓
Calibration offset: -1
①
Press the ENTER key. The setting is decided
and the function menu selection screen is
displayed.
Offset is adjustable within the range from +3.0℃ to
-3.0℃.
Temperature correction in
overheating prevention device:
The temperature of overheating prevention device is
automatically corrected by the same amount when
that of controller is collected.