Operation method, Input program – Yamato Scientific IN603W Low Temperature Incubators User Manual

Page 29

Advertising
background image

25

Operation Method

Input Program

1

Select program function
• Check that the power in turned on.

PROGRAM

TROUBLE

ENTER

CANCEL

Select step

S01

Program OPR

Press the PROGRAM key. The program menu

starts and step number selection screen is
displayed.
Press the "▼▲". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.

™

The "S01" is displayed when no program is

registered. In this case, the "▼▲" are invalid.

Press the ENTER key. The selected step

number is decided and the setting item selection
screen is displayed. The "Set TEMP" is
displayed first.

Press the CANCEL key to cancel the program

menu.

2

Edit step
• The following four setting items are included in a step.


• The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different.

It enters into "Wait" if the temperature does not reach to the setting temperature within a setting
time.

Set Temp.

TEMP

.

TIME

Set Time

TEMP

.

START

.

(Operation with fixed rate inclination)

Setting item

Setting range

Setting temperature

Setting temperature range by product type

Setting time

0 minute to 999 hour and 59 minutes, End

Repeat initiating step

None, Registered step numbers 1 to 32

Repeat count

1 to 9999, Infinity

Advertising