Trouble shooting – Yamato Scientific IN603W Low Temperature Incubators User Manual

Page 60

Advertising
background image

56

In the Event of Failure…

Trouble Shooting

Problem Possible

Cause

Solution

Earth leakage breaker failure

Replace the part.

Power switch failure

Replace the part.

The device does not start
when turning on the power
switch.

Power source failure

Connect to the appropriate
power source

Heater disconnection

Replace the part.

SSR failure

Replace the part.

Temperature does not rise.

Temperature controller failure

Replace the part.

Temperature sensor failure

Replace the part.

Temperature controller failure

Replace the part.

Clogging of condenser with dust

Clean the fin on condenser

Much frost on evaporator

Defrost

Relay failure

Replace the part.

Power source failure

Connect to the appropriate
power source

Temperature does not fall.

Refrigerator failure

Repair or replace the part.

SSR failure

Replace the part.

Heater does not stop working
when the temperature reaches
setting value.

Temperature controller failure

Replace the part.

If power failure occurs…

The unit returns automatically to start operation automatically with the same condition as just
before the failure when it occurs during operation and is recovered. When I used a backup

power supply for blackout return, but a blackout occurred for need in less than 30 minutes more
than 30 minutes as for the charge time of the backup power supply since I turned on electricity

in a device, there is the thing that a blackout does not return temporarily. The blackout return

function operates normally after it was charged a backup power supply enough because I am not
abnormal.