Before Using This Unit
Requirements for Installation
1. Choose a proper place for installation
• Do not install this unit in a place where:
♦ Rough or dirty surface.
♦ Flammable gas or corrosive gas is generated.
♦ Ambient temperature 35°C and above or 5°C and below.
♦ Ambient temperature fluctuates violently.
♦ There is direct sunlight.
♦ There is excessive humidity and dust.
♦ There is a constant vibration.
♦ Not
horizontal
surface.
♦ The power source is instable.
• Install this unit on a stable place with the space as shown below.
30cm or more
15cm
or
more
15cm or more
15cm
or
more
Back side
Front side
2. Installation on horizontal surface
• Use the equipment on the horizontal and firm place to
keep the water inside the chamber horizontal. If the
equipment tilts and the heater appears from the water
surface, the temperature on the area above the water
rises and a heater failure or operation stop due to
water level detector function may occur.
• The weight of main unit is approximately 65 to 95kg.
Carry and install the equipment carefully by two or
more persons.