Before using this unit, Requirements for installation – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual

Page 9

Advertising
background image

6

Before Using This Unit

Requirements for Installation

1. Choose a proper place for installation

• Do not install this unit in a place where:

♦ Rough or dirty surface.

♦ Flammable gas or corrosive gas is generated.
♦ Ambient temperature 35°C and above or 5°C and below.

♦ Ambient temperature fluctuates violently.

♦ There is direct sunlight.

♦ There is excessive humidity and dust.

♦ There is a constant vibration.

♦ Not

horizontal

surface.

♦ The power source is instable.

• Install this unit on a stable place with the space as shown below.

30cm or more

15cm

or

more

15cm or more

15cm

or

more

Back side

Front side

2. Installation on horizontal surface

• Use the equipment on the horizontal and firm place to

keep the water inside the chamber horizontal. If the
equipment tilts and the heater appears from the water
surface, the temperature on the area above the water
rises and a heater failure or operation stop due to
water level detector function may occur.

• The weight of main unit is approximately 65 to 95kg.

Carry and install the equipment carefully by two or
more persons.

Advertising