Identification of Parts
Operation System
(3)Collection of distilled water
● For about 11mim. from the power switch pushed, distilled water condensed in
condenser is
stored into
distilled water tank, after initial water is drained by
solenoid valve (to drain the initial
distilled water) opened. The unit judges water full and stops distillation when the highest
float
switch inside the tank is activated. After collecting distilled water and using some amount of distilled
water ,the distilled water is made automatically.
Distilled water stored is taken out by
pump(to take distilled water) through
solenoid valve to
take distilled water and
distilled water outlet.
(4)Collection of ion-exchanged water
●Ion-exchanged water is taken out by through
decompressor,
solenoid valve for raw water ,
pre-treatment cartridge ,
Ion-exchange resin cartridge(CPC-N),
electrode for ion-
exchanged water ,
solenoid valve to take ion-exchanged water , and
Ion-exchanged water
outlet .