Maintenance

Exchange Heaters

● If the heater snapped or was broken, exchange the heater as follows.(Refer to P.29∼32 for

exchanging)

1. Turn “OFF” the circuit breaker.
2. Close the faucet.
3. About more than 30 min. later after turned “OFF” the circuit breaker, open the front door of this unit and

the valve to drain boiler water.

4. Open the left side board, loosen the four connecting screws in right side of terminal rack, and pull the

heater terminal.

5. Pull the heater terminal from the grommet.
6. Remove the cap nut of the heater, and pull the heater.
7. Remove the packing and the cap nut from the broken heater.
8. Install the packing and the cap nut to the new heater.

At this time, do not touch it with empty hands not to apply the finger marks.

9. Insert it into the boiler with the heater mark “YK-W-3” upward.

10. Insert the heater terminal into the grommet, and fix it in the terminal rack with checking the position of

installing heater lead.

11. Install the left side board on the unit.
12. Close the valve to drain boiler water.
13. Open the faucet after closing the front door.
14. Turn “ON” the circuit breaker.
15. Push the power switch in the operation panel, and the supply of boiler water starts.