OPERATING INSTRUCTIONS___________________________________________________________________________































8. If the load current exceeds the preset current limit, the

CONSTANT CURRENT indicator will light. In this case, the
power supply automatically switches to the constant current
mode and further rotation of the VOLTAGE control will not
increase the output voltage.


Setting Current Limit
1. Determine the maximum safe current for the device to be

powered.

2. Temporarily short the (+) and (-) terminals of the power

supply together with a test lead.

3. Rotate the VOLTAGE control away from zero sufficiently

for the Constant Current indicator to light.

4. Adjust the CURRENT control for the desired current limit.

Read the current value on the mA meter.

5. The current limit (overload protection) has now been

preset. Do not change the CURRENT control setting after
this step.

6. Remove the short between the (+) and (-) terminals and

hook up for constant voltage operation.


Typical Constant Current Operation
1. Before connecting the device to be powered to the power

supply, determine the maximum safe voltage to be applied,
set the A/B Meter s election switch to the A (up) position, and
set the VOLTAGE control to obtain that voltage reading on the
Volt meter.

2. Determine the desired constant current value.
3. Set the CURRENT control to minimum (fully

counterclockwise).

Figure 4. Typical constant voltage operation

4. Turn off the power supply and connect it to the device to

be powered.

5. Turn on the power supply. The CONSTANT CURRENT

indicator should light.

6. The current can be read directly on the mA meter.