Calibration checks, Seup test, Figure 9 calibration check setup – Bird Technologies 4304A User Manual

Page 32: Calibrating a bird element 4304a-1, Seup test calibrating a bird element 4304a-1

Advertising
background image

21

Calibration Checks

Seup Test

Note: This test determines if the wattmeter needs recalibration:

1.

The following equipment is required, connected in series with the Bird
4304A (Refer to Figure 9).

RF source

Low pass filter

Reference, known accurate, wattmeter

50

Ω load

Cables

Note: Keep all connections as short as possible to minimize mis-
match and insertion losses.

2.

Set the RF source to produce a signal with power in the upper one-third of
the scale for each power range.

3.

Compare the reading on the Bird 4304A with the reference wattmeter.

Note: If the unit is outside the stated accuracy range, return it to a
Bird Service Center for calibration.

Figure 9 Calibration Check Setup

Calibrating a Bird Element 4304A-1

Required Equipment

RF source capable of 100W from 25 MHz to 1000 MHz

RF source capable of 500 W at 405 MHz

Power standard

Model 4304A Wattmeter

5

15

50

150

500

Low Pass

Filter

4304A

RF Source

50 Load

Ω

Reference

Wattmeter