Setup, Dip switch – Bird Technologies 4421 User Manual

Page 38

Advertising
background image

Bird 4421 RF Power Meter

26

Setup

DIP Switch

Set the DIP switches according to application needs and the requirements of
the controller. Available settings and factory defaults are listed in Figure 12.



Note:

There are several button styles on DIP switches (slide, rocker,

lever). Examine the DIP switch to determine the ON and OFF positions.

Figure 12 RS-232 Interface Default DIP Switch Settings



Note:

If you change switch settings when the unit is ON, the new

settings will not become effective unitl you turn power OFF and then
back ON.

Switch

Function

Description

1

Stop Bit

ON 1 Stop Bit
OFF2 Stop Bits

*

* Default factory setting

2

Command
Mode

ON ENT command needed before sending reading to
controller*
OFFTrigger automatically sends reading
Reading also sent when front panel push button pressed

3

Word Length

ON 8 Data Bits*
OFF7 Data Bits

4,5

Parity

(4) (5)
ON ON
ON OFF
OFFON
OFFOFF

No Parity*
Odd Parity
Even Parity
Mark Parity

6,7,8

Baud Rate

(6) (7) (8)
ON ON

ON

ON OFFON
ON OFFOFF
OFFON ON
OFFON OFF
OFFOFFON
OFFOFFOFF

Auto Baud

300
600
1200
2400*
4800
9600

† For more information see “Auto Baud”

2

1

4

3

6

5

8

7

ON