6 Methods / Parameters

780/781 pH/Ion Meter, Manual

65

6.3

Temperature measurement (T mode)

6.3.1

Measuring parameters

PARAMETER T

├

measuring parameters

: ├

electr.id

├

drift

├

method

├

delta measurement

└

stirrer

The measuring parameters describe the settings of direct importance

for the measurement.

electr.id

12 ASCII characters

You can assign any name you want to the connected temperature sen-

sor. This sensor Id is then contained in the report header for measured

values.

drift

0.5...1.0...999.9 °C/min, OFF

A measuring signal which changes as a function of time is regarded as

being sufficiently constant when its drift drops below the value defined

here. This can be used as criterion for the pH/Ion Meter if a displayed

measured value is to be stored or documented by printing a report (see

print criterion and storage criterion in Section 5.2 and Section 5.3).

Please note that for a drift-controlled measured value recording this drift

parameter must not be set to '

OFF

', as otherwise no result will be ex-

ported.
If this drift condition is not fulfilled then this is indicated by two constant-

ly blinking triangles in the display.
An active drift control increases the reproducibility of the measurements

as a constant condition is used to define the measured result. A smaller

value for the measuring parameter Drift requires a more stable measur-

ing signal than a larger one. The set measuring parameter is also pro-

portional to the measuring uncertainty of the measured value obtained.

method

8 ASCII characters

All parameter settings can be stored together or loaded as a method

(see Section 6.1). The name of the current method is then given here.

This field is only used for information here and cannot be edited direct-

ly.

delta measurement: ON, OFF
The delta measurement permits the display of the measured value rela-

tive to a freely selectable constant reference value:

reference

-999.9...0.0...999.9 °C

Advertising